Parameter Typical Return
0|OFF|1|ON,
*RST 1 step with a valueof OFF
<list value 1>,
<list value 2>,
<list value 3>...
To generate triggers at the beginning of the second step of a 3-step list:
LIST:TOUT:BOST OFF,ON,OFF, (@1)
l The order in which the list values are entered determines the sequence when the list executes.
l To create a valid list, the Voltage, Current, BOST, EOST, and Dwell lists must all be the same
length or have a length of 1, which is interpreted as having the same length as the longest list.
l This command overwrites any previously programmed list; it does not append to the previous list.
Keysight N6705C Operating and Service Guide 299
7 SCPI Programming Reference