[SOURce:]VOLTage:SLEW:MAXimum 0|OFF|1|ON, (@<chanlist>)
[SOURce:]VOLTage:SLEW:MAXimum? (@<chanlist>)
Enables/disables the maximum slew rate override. When enabled, the slew rate is set to its maximum
value. When disabled, the slew rate is set to the immediate value set by the
[SOURce:]VOLTage:SLEW command. Use [SOURce:]VOLTage:SLEW? MAX to query the maximum
slew rate.
Parameter Typical Return
0|OFF|1|ON, *RST OFF 0or1
Enable the maximum slewrateoverride: VOLT:SLEW:MAX ON, (@1)
l The [SOURce:]VOLTage:SLEW:MAX command is coupled to the [SOURce:]VOLTage:SLEW com-
mand. If [SOURce:]VOLTage:SLEW sets the rate to MAX or INFinity,
[SOURce:]VOLTage:SLEW:MAX is enabled. If the slew rate is set to any other value,
[SOURce:]VOLTage:SLEW:MAX is disabled.
7 SCPI Programming Reference
348 Keysight N6705C Operating and Service Guide

Table of Contents