Test Record Form - Keysight N6783A-BAT and N6783A-MFG
N6783A-BAT, '83A-MFG Report Number ______ Date__________
Test Description Model Min. Specs Results Max. Specs
Constant Voltage Tests
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
N6783A-BAT
N6783A-MFG
N6783A-BAT
N6783A-MFG
− 5 mV
Vout − 5 mV
7.982 V
5.984 V
Vout − 9 mV
Vout − 8 mV
_______
_______
_______
_______
_______
_______
+ 25 mV
Vout + 5 mV
8.018 V
6.016 V
Vout + 9 mV
Vout +8 mV
CV Load Effect: Both – 6 mV _______ + 6 mV
CV Source Effect: Both – 2 mV _______ + 2 mV
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
+ 8 mV
+ 1.5 mV
Transient response 6 V
Voltage @ 45 μs:
Voltage @ 75 μs with Option 761:
N6732B
N6742B
– 75 mV
– 90 mV
_______
_______
+ 75 mV
+ 90 mV
Constant Current Tests
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Low current measured over interface:
Both
Both
Both
Both
Both
+ 3.2 mA
Iout − 0.6 mA
2.9952 A
Iout − 3.6 mA
Iout − 225 µA
_______
_______
_______
_______
_______
+ 6.8 mA
Iout + 0.6 mA
3.0048 A
Iout + 3.6 mA
Iout + 225 µA
CC Load Effect: Both – 2 mA _______ + 2 mA
CC Source Effect: Both – 1 mA _______ + 1 mA
Negative Current Output: Both – 1.9942 A _______ + 2.0058 A
Test Description N6783A-BAT Setting N6783A-MFG Setting
Voltage Programming & Readback Min Voltage:
Voltage Programming & Readback High Voltage:
CV Load Effect, Source Effect, Ripple and Noise:
Transient Response:
Current Programming & Readback Min Current:
Current Programming & Readback High Current:
Current Readback Low Current:
CC Load Effect, Source Effect:
Negative current output:
15 mV,3 A
8 V, 3 A
8 V, 3 A
6 V, from 0.15 A to 1.5 A
5 mA, 8 V
3 A, 8 V
150 mA, 8 V
3 A, 8 V
0 V, – 2 A
15 mV,3 A
6 V, 3 A
6 V, 3 A
6 V, from 0.15 A to 1.5 A
40 mA, 6 V
3, 6 V
150 mA, 6 V
3 A, 6 V
0 V, – 2 A
Keysight N6705C Operating and Service Guide 433
8 Verification and Calibration