EasyManua.ls Logo

Keysight N6705C - Page 45

Keysight N6705C
476 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Characteristic Keysight N6705C
Digital Control Characteristics
Maximum voltageratings
Pins 1 and 2 as FLT output
Pins 1 - 7 as digital/trigger
outputs (pin 8 = common)
Pins 1 - 7 as digital/trigger
inputs and pin 3 as INH input
(pin 8 = common)
+16.5 VDC/-5 VDC between pins
(pin 8 is internally connected to chassis ground).
Maximum low-level output voltage = 0.5 V @ 4 mA
Maximum low-level sink current = 4 mA
Typical high-level leakage current = 1 mA @ 16.5 VDC
Maximum low-level output voltage = 0.5 V @ 4 mA;
1 V @ 50 mA; 1.75 V @ 100 mA
Maximum low-level sink current = 100 mA
Typical high-level leakage current = 0.8 mA @ 16.5 VDC
Maximum low-level input voltage= 0.8 V
Minimum high-level input voltage= 2 V
Typical low-level current = 2 mA @ 0 V (internal 2.2k pull-up)
Typical high-level leakage current = 0.12 mA @ 16.5 VDC
Interface Capabilities
LXI Core 2011:
USB 2.0 (USB-TMC488):
10/100/1000 LAN:
Built-in Web server:
GPIB:
10/100/1000 Base-T Ethernet (Sockets, VXI-11 protocol, Web interface)
Requires KeysightIO Library version M.01.01 or 14.0 and up
Requires KeysightIO Library version L.01.01 or 14.0 and up
Requires a Web browser
SCPI - 1993, IEEE 488.2 compliant interface
Regulatory Compliance:
EMC:
Safety:
Complies with European EMC Directive for test and measurement products:
IEC/EN 61326-1; CISPR 11, Group 1, class A; AS/NZS CISPR 11; ICES/NMB-001
Complies with Australian standard and carries C-Tick mark
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Complies with European Low Voltage Directive and carries the CE-marking.
Conforms to USand Canadian safety regulations.
Environmental
Operating environment:
Temperature range:
Relative humidity:
Altitude:
Storagetemperature:
Indoor use, installation category II (for AC input), pollution degree2
0°C to 55°C (output current is derated 1% per °C above40°C
Up to 95% (non-condensing)
Up to 2000 meters
-30°C to 70°C
Acoustic Noise Declaration
This statementis provided to
comply with the requirements
of the German Sound Emission
Directive, from 18 January 1991
Sound Pressure Lp <70 dB(A), At Operator Position, Normal Operation, According to EN 27779
(Type Test).
Schalldruckpegel Lp <70 dB(A), Am Arbeitsplatz, Normaler Betrieb, Nach EN 27779 (Typ-
prüfung).
OutputTerminal Isolation: No output terminal may bemore than 240 VDC from any other terminal or chassis ground.
N6781A/N6785A Note:
When using the AUX measurement input terminals on Models N6781A and N6785A, no outputor
input terminal may bemore than ±60 VDC from any other terminal or chassis ground.
Keysight N6705C Operating and Service Guide 45
1 Quick Reference

Table of Contents

Related product manuals