Keysight Technologies N9040B UXA Signal Analyzer Service Guide 123
Instrument Messages
Condition Messages
207 2 Burst Not Found E The burst signal cannot be detected
because of inappropriate parameter settings
or incorrect signal.
An in appropriate parameter setting could
cause the signal to be partially, rather than
fully, on the display, Burst Search Threshold
and/or Burst Search Length may need to be
adjusted.
An incorrect signal could have either
insufficient power, the rising or falling edges
cannot be detected, or the burst is less than
126 microseconds.
Carrier signal is not actually bursted.
W-CDMA: Either the signal being analyzed
has insufficient power, the rising or falling
edges cannot be detected, or the burst is
less than 126 microseconds.
W_CDMA: Cannot synchronize
measurement with PRACH channel for
Power Control measurement, because the
signal cannot be found. Make sure PRACH is
present in the W-CDMA uplink signal, and
that the preamble signature and scramble
code are set correctly.
GSM: Data was acquired but a GSM burst
was not found, with the timeslot mode
disabled.
NADC, PDC: A valid burst is not found when
the Device is MS.
1xEV-DO: Data was acquired but a 1xEV
burst was not found, with the timeslot mode
disabled.
Bluetooth: The burst that has been found
does not correspond to the currently
selected Bluetooth packet type (the burst
length may be too short).
WLAN: The instrument cannot find a valid
WLAN burst. You may need to extend the
search length.
In the PSA, this error was reported as one of
the following error numbers: 10772, 13104,
10160, 10286, 10420, 10454, 10614,
10904, 10928, 13074, 10287
Err# Bit in
Status
Register
Message Error or
Warning
More Information