124 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Condition Messages
207 2 Burst not found;with
selected Time Slot
E The selected timeslot does not contain the
expected burst.
209 3 Timing Error E
209 3 Timing Error:No time
ref pilot burst
E The pilot burst used for time reference is not
active.
211 4 Carrier(s) incorrect
or missing
E In the PSA, this error was reported as one of
the following error numbers: 10165, 10173,
10178, 10419, 10421, 10535, 10560,
10642, 10648, 10650, 10960
213 5 Freq Out of Range E
213 5 Freq Out of Range;
System input (IF)
E One or more system input frequencies are
out of range. If using a frequency list, check
that all entries are valid for current
measurement mode.
213 5 Freq Out of Range;
External LO
E One or more external LO frequencies are out
of range. Check that the LO frequency limits
are set correctly and check the entered
measurement frequencies and
measurement mode.
215 6 Sync Error E W-CDMA: Cannot sync DPCCH pilot.
Cannot synchronize measurement with
DPCCH pilot for Power Control
measurement, because the pilot signal
cannot be found. Make sure DPCCH is
present in the W-CDMA uplink signal, and
that the slot format and scramble code are
set correctly.
215 6 Sync Error;No pilot
burst
E There is no Pilot burst detected.
215 6 Sync Error;Sync code
not found
E Synchronization code is not found in the
measured time slot.
215 6 Sync Error;No freq
ref pilot burst
E The pilot burst used for frequency reference
is not active.
215 6 Sync Error;Midamble
sync fail
E Failed to find the uplink slot, which caused
the synchronization with the midamble to
fail.
215 6 Sync Error;Preamble
length zero
E Burst type is “Data” or “Preamble” and the
measurement cannot find a Preamble
Err# Bit in
Status
Register
Message Error or
Warning
More Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals