174 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
5. Place the RF front end troubleshooting board on an ESD safe bag or foam
to ensure nothing shorts out. See Figure 5-4.
Figure 5-4 RF Front End Troubleshooting Board Placement
6. Turn the instrument on and allow it to complete its full boot up process to
the Spectrum Analyzer application.
7. Turn Auto Align off by pressing System, Alignments, Auto Align, choose
Off from the drop-down menu.
You may notice an RF Alignment failure message on the analyzer display. This will occur when
the test board is connected.
Resistors on the test board can get very hot. Handle with care.