194 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A2 Analog IF Troubleshooting
10.Connect the W42 cable to the MMCX female to SMA female connector.
Use an appropriate cable to go from the SMA connector to the RF input of
a functioning spectrum analyzer to verify the 22.5 MHz I.F. and amplitude
is correct.
11.Press Freq, 22.5 MHz, Span, 1 MHz, Peak Search on the functioning
spectrum analyzer.
12.The analyzer marker should read 22.5 MHz at 25 dBm ± 4 dB as shown in
Figure 6-4.
Figure 6-4 A2 Analog IF Output
13.If the 22.5 MHz signal is not measuring the correct power level, do not
assume the Analog IF is the most probable cause until the 3rd LO
frequency and power level have been verified below.
14.If the 22.5 MHz signal is within tolerance, carefully reconnect the W42
cable to A3J19.
You should hear a distinct snap when reconnecting the W42 cable. If this cable is not installed
properly, intermittent signal fluctuations may occur on the analyzer display.