EasyManua.ls Logo

Keysight N9040B Option 508 - Page 380

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
380 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area - Options 508, 513, 526
7. Refer to Figure 15-27. Place the front brace (1) into the chassis and
secure with the eleven screws (2) (0515-0372). Torque to 9 inch-lbs.
Figure 15-27 Front Brace Replacement
8. Replace the RF bracket, front panel, and instrument outer case as
described on page 360.