EasyManua.ls Logo

Keysight N9040B Option 508 - Page 408

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
408 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
RF Area - Options 544, 550
4. Refer to Figure 15-56. Unplug ribbon cables W22 and W28, and wire
harness W12 from the Front End Controller board.
Figure 15-56 Front End Controller Board Cables
5. Refer to Figure 15-57. From the bottom of the instrument, remove the six
screws (2) (0515-0372) that attach the Front End assembly (1) to the
chassis. The Front End assembly can now be removed.
Figure 15-57 Front End Assembly Bottom Screws - Removal