434 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
A28 Wideband Extension Board (Option RTS)
4. Serial Number Prefix ≥ 5605:
— Refer to Figure 15-82. Remove the RF bracket (1) by removing the
thirteen screws (2) (0515-0372), and the six screws (3)
(0515-1946), using the T-10 driver.
Figure 15-82 RF Bracket Removal,
Serial Number ≥ Prefix 5605