506 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Functional Tests
Test Equipment
Cables
All 3.5 mm (1 meter) Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
11500E 8120-4921
All Cable, BNC
(3 required)
120 cm (48 in.) BNC cable 10503A
Signal Source
All Synthesized Sweeper
Frequency: 10 MHz to 50 GHz
a
Harmonic level: < - 40 dBc
Amplitude range: 10 to - 20 dBm
Frequency Accuracy: 0.02%
PSG
b
83630B,
83640B,
83650B
Power Meter
All Power Meter Power Reference Accuracy: ±1.2%
Compatible with power sensor
E4418B E4419B
All Power Sensor Frequency Range: 50 MHz to 3.66 GHz
Amplitude Range: -70 to –20 dBm
8481D 8487D
8485D
508, 513 Power Sensor Frequency Range: 50 MHz to
13.66 GHz
Amplitude Range: -30 to +20 dBm
8481A
526 Power Sensor Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: -30 to +20 dBm
8485A
Preamp Low Power Sensor
Frequency Range: 50 MHz to 50 GHz
c
Amplitude Range: –70 to –20 dBm
8487D
8485D
c
Miscellaneous Equipment
All Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
0955-0306
526 Power Splitter, 3.5 mm Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667B
508, 513 Power Splitter Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667A 11667B
Analyzer
Option
Item Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model

Table of Contents