EasyManuals Logo

Labconco Logic+ A2 User Manual

Labconco Logic+ A2
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
139
12. Push the tube into the fitting until it is properly seated. The tube will go
approximately ¾ inch (19 mm) into the fitting when properly seated.
13. Tighten the 90 degree fitting compression nut hand-tight and then, using a 7/16-
inch wrench, tighten it at least ¾ turn more.
14. Close the service valve in the cabinet and then slowly open the shutoff valve on
the service valve. Test all fittings for leakage. Tighten the tube nut slightly if
needed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Labconco Logic+ A2 and is the answer not in the manual?

Labconco Logic+ A2 Specifications

General IconGeneral
BrandLabconco
ModelLogic+ A2
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals