EasyManua.ls Logo

Labconco Logic+ A2 - Page 39

Labconco Logic+ A2
201 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39
10. If a Type C1 model, and B-Mode operation was selected previously, the Active
Protection Screen will appear. This allows the selection of the time for which the
Type C1 will continue to operate (while sounding an alarm) when a remote
(building) exhaust failure occurs. This allows the user time to secure critical work
and exit before the cabinet shuts off. Set in 10 second increments up to 5
minutes (300 seconds). Use [UP] or [DOWN] to select the Active Protection
time. Press [OK/MUTE].
C1
B MODE
ACTIVE PROTECTION
10 SECONDS

Table of Contents

Related product manuals