38
6. On a Type A2 or C1 only, if field installing, or removing the Airflow Sensor,
change the selection accordingly, press [OK/MUTE].
7. On all biosafety cabinet types, if field installing, or removing the UV Light, change
the selection accordingly (screen shown above), press [OK/MUTE].
8. On Type A2 models only, if field installing, or removing the Canopy connection,
change the selection accordingly, press [OK/MUTE].
Note: When a Type C is selected previously, the Canopy Screen will not appear, but
the software will default the selection correctly based on whether A-mode (No Canopy)
or B-mode (Canopy) was previously selected.
9. The Sash Height selection will display next, and should NOT be changed, unless
the model is a C1, and the user has decided to switch between 8-inch and 10-
inch sash operation. Press [OK/MUTE].
If the biosafety cabinet is a Type A2 or B2, this is the final screen selection. If a C1,
proceed to step #10.