6-24
Using the 10 and 20 cm Survey Chambers
Disable 'CRC' to remove CRC from the data. Note that the LI-8100APP does
not use CRC checking.
Note that when the LI-8100A outputs data, each field is "marked up" using
eXtensible Markup Language (XML) to delimit that field. For example, when a
CO
2
value is output, the data value is placed between two "tags" that describe
what that value is, as in:
<CO2>350.21</CO2>
Enable 'Strip' to remove the markup from the data stream. The resulting data
set is a data stream where each data field is separated by a space.
TCP/IP Output Fields
Tap on the TCP/IP field to configure the data values to be sent via the TCP/IP
port, if desired. Note that you can configure the TCP/IP (wireless or wired)
and/or RS-232 options while the other is enabled; changes will not take effect
until data logging is stopped and restarted, however. For example, if you are
currently logging data to the PC, you can configure the wireless output, but
you must first stop logging data to the PC before beginning wireless output
with the new configuration.
The data values that can be output are as shown above at RS-232 Output,
with the addition of GPS values (latitude, longitude, GPS status, speed, and
course).
5. Click on the Setup menu and choose Start Measurement. The Start
Measurement screen appears (below).