2-14
Initial Setup
7. There are 10 EPDM type plugs that can be inserted into unused glands on the
Auxiliary Sensor Interface; the plugs prevent water, insects, dirt, etc. from
entering the interface box. Remove the top cover and insert the narrow end of
the plug through the back of the gland and tighten the cap(s) (below). The
plugs should always be inserted when there are glands that do not have wires
inserted through them.
8. Note too that there is a length of Santoprene tubing in the Auxiliary Sensor
Interface spares kit. This tubing can be cut to length and used for small gauge
wires that may not be able to be compressed sufficiently with the glands. It
can also be used for oddly shaped wires that can be difficult to seal with the
gland caps.