Using the 10 and 20 cm Survey Chambers
6-13
The Repeat Measurement page allows you to repeat the defined protocol at a
regular clock interval. These functions are particularly useful when making
long term, unattended measurements.
For example, you could specify a 90 second Obs. Length, 45 second
Deadband, Obs. Delay of 2 minutes, and Obs. Count of 3. This protocol
could then be repeated every hour for 240 hours (10 days). The resulting data
set would include 240 measurements, with each measurement consisting of
three 90-second observations on the chosen port. The maximum number of
repeats is 12000 (Set To Max button).
You can elect to turn off the flow pump between repeated measurements to
conserve battery life and extend the life of the pump by enabling the check
box.
5. Choose Start Measurement from the Setup menu to open the Start a
Measurement dialog.