Root NIBPT Chapter 3: Operation
www.masimo.com 63 Masimo
Calibration for NIBP
The Calibration option on the NIBP menu allows a qualified service professional to access
calibration settings and tools for the NIBP module. For more information, see NIBP Module
Calibration Test on page 150.
Note: This section is provided as a reference and intended for qualified service professionals
only.
Sounds
Use the Sounds screen to control the volume level of sounds and duration of audio pause for
Root.
Option Description Factory
Default
Setting
Configurable Settings
Alarm
Volume
Sets the alarm volume level.
Highest
volume
Slide towards the left to decrease
volume to silence.
Pulse Tone
Volume
Sets the pulse tone volume
level.
3
Slide towards the left to decrease
volume to silence.
Audio Pause
Duration
Sets the length of time that
the audible alarm remains
silenced, when Audio Pause
is enabled. See Audio Pause
on page 80.
2 minutes
1, 2, 3 minutes, Permanent*,
Permanent with Reminder*.
If Permanent is selected, there will
be no audible alarms when Audio
Pause is enabled, but visual
alarms will still display.
If Permanent with Reminder is
selected, a tone will sound every
three (3) minutes as a reminder
that Permanent is active when
Audio Pause is enabled.
SmartTone
Allows the audible pulse to
continue to beep when the
pleth graph shows signs of
motion.
Off On or Off
*Requires the all mute enabled option to be toggled to ON in the Access Control menu. See
Access Control on page 70.