Operating Instructions - CPC6050 101
Modular Pressure Controller
CPC6050
:STABle? Returns 1 if in Control mode and pressure is
stable, 0 otherwise
:STATe ON | OFF | 1 | 0 ON or 1 puts in Control mode , OFF or 0 puts in
Measure mode
[:STATe]? Returns 1 if in Control mode, 0 otherwise
SENSe
[:PRESsure][R]
:BARometer? Returns the current barometer reading in active
channel’s units
:CORRection
:OFFSet? Returns the tare value in current units
:OFFSet <f> Set the head pressure correction value in current
units
:OFFSet
:STATe? Returns 1 if the tare value is currently enabled, 0
if not.
:STATe ON | OFF | 1 | 0
:HEAD? Returns the gas type, current head pressure cor-
rection in current units for the active channel. Ex//
SENS:PRES:CORR:HEAD?
:SENS:PRES:CORR:HEAD DRY AIR,
0.000000
:HEAD <”CCC”, f> Set the gas type and head pressure correction
value in current units. Requires the gas type(CCC
= AIR, HELIUM, NITROGEN) in double quotes,
numeric value (oat)
:STATe? Returns 1 if the head pressure correction is cur-
rently enabled, 0 if not.
:STATe ON | OFF | 1 | 0 Enable or disable the head pressure correction
:VOLume? Returns the calculated volume by the regulator
in cubic centimeters. This value is constantly
changing while in control mode
:VOLume
:VALue? Returns the regulator control volume. May be
AUTO or a number in cubic centimeters
:VALue <f> Set the regulator control volume to AUTO or a
number (0 to 1000 for a pump, 0 to 2000 for a
SVR)
:FILTer
[:LPASs]
:STATe? Returns if the lter is enabled (0 is o, 1 is on)