102 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
:STATe ON | OFF | 1 | 0 Set the lter enable
:BAND? Returns the lter window
:BAND <f> Set the lter window
:FREQuency? Returns the lter frequency
:FREQuency <f> Set the lter frequency
:INLimits? Returns current pressure reading (oat), stable
ag (0 not stable, 1 stable)
:INLimits
:TIME? Returns the stable delay in seconds
:TIME<f> Set the stable delay in seconds
:RANGe? Returns primary sensor maximum pressure,
units, and pressure type, i.e. “1000.000psig”
:RESolution? Returns sensor resolution (integer 4 to 6)
:RESolution <n> Set the sensor resolution
:SLEW? Returns the rate in current units
SOURce
[:PRESsure]
[:LEVel]
[:IMMediate]
[:AMPLitude]? Returns the setpoint in current units
[:AMPLitude] <f> Sets the setpoint in current units
:VENT? Returns 1 if in vent mode and pressure is not fully
vented (stable) yet, 0 otherwise. Proper usage:
put in vent mode, then query this value until it is 0
again to know the system is vented.
:VENT ON | OFF | 1 | 0 ON or 1 puts in vent mode , OFF or 0 puts in
Measure mode
:SLEW? Returns the rate setpoint in current units
:SLEW <f> Sets the rate setpoint in current units
:SLEW
:MODe? Returns MAX when highspeed mode is active.
Returns LIN otherwise.
:MODe <CCC> Set the control mode by putting in either High-
speed or Custom mode. Note: the overshoot ag
and control mode are not independent in Mensor
products.
:OVERshoot? Returns 1 if the channel’s control behavior is >50.
Returns 0 if the control behavior <=50.