Operating Instructions - CPC6050 75
Modular Pressure Controller
CPC6050
Alarm_Limit? <sp>xxxxxxx<cr><lf> Returns the Alarm Limit
Asset_tag 16 char string General purpose string for customer
use
Asset_tag? <sp>ssssssssssssssss<cr><lf> Return customer asset tag string
Autorange ON or OFF Sets whether the autorange function
is enabled or disabled
Autorange? <sp>(ON or OFF)<cr><lf> Returns whether the autorange func-
tion is enabled or disabled.
Autozero none Re-zero all the ranges. These adjust-
ments are not password protected
and are saved permanently through
power cycles while the instrument
is in native mode (temporary while
in emulation mode). This command
takes approximately 60 seconds but
may take longer depending on the
time to become stable.
Autozero? S,T,X,X Returns autozero data where S
represents state (0 = complete, 1 =
local autozero, 2 = remote autozero),
T represents the estimated time to
complete in seconds, and x is a (0)
character since this data location is
not used at this time.
Autozeroabort none Aborts autozero. Any transducers that
have been zeroed will not revert to
previous zero osets.
Auxdisp <n> NONE,PEAK,RATE,RATE SETPOINT,
UNCERTAINTY,UNITS,BAROMETER
Sets the auxiliary display setting for
index <n> (0 to 2). If <n> is left out,
defaults to index 0.
Auxdisp? <n> <sp>NONE,PEAK,RATE,RATE
SETPOINT,UNCERTAINTY,UNITS,
BAROMETER <cr><lf>
Returns the auxiliary display setting
for index <n>. If <n> is left out, de-
faults to index 0.
B? <sp>+n.nnnnnE+nn<cr><lf> Returns the B channel pressure read-
ing in current units
BR? <sp>+n.nnnnnE+nn<cr><lf> Returns the B channel rate reading in
current units
BRS? <sp>{YES or NO}<cr><lf> Returns the B channel rate stable ag
BS? <sp>{YES or NO}<cr><lf> Returns the B channel pressure
stable ag
Baro? <sp>+n.nnnnnE+nn<cr><lf> Returns reading from barometric
transducer or “NO BAROMETER” if
one isn’t installed
Barounits units code or text in table below Sets the barometer pressure units
Barounits? <sp>CCCC<cr><lf> Returns the barometer units in a text
string