78 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
DIO 2 or 0 2 turns on the rst digital output pin, 0
turns it o
DIO? <sp>n<cr><lf> Returns status of the rst input and
output pin. Bit0 = input’s status, Bit1 =
output’s status.
DOUTFUNC <n><sp><CCCCCC><cr><lf> Sets the function for output pin <n> to
NONE, STABLE, PUMP, MEASURE,
CONTROL, VENT
DOUTFUNC? <n> <sp><CCCCCC><cr><lf> Returns the function for output pin
<n>
DINFUNC <n><sp><CCCCCC> Sets the function for input pin <n>
to NONE, MEASURE, CONTROL,
VENT, KEYLOCK, START
DINFUNC? <n> <sp><CCCCCC><cr><lf> Returns the function for output pin
<n>
DOUTSTATE <n><sp><HIGH/LOW/1/0> Turns output pin <n> to high, or low.
Sets the function for that pin to “None”
DIOSTATE? <sp><n><cr><lf> Bit 0-2 is the status of the Input bits,
Bit 3-5 are the status of the Output
bits. Returns an integer between 0
and 63
DOC mm/dd/yyyy Sets the date of cal for the active
transducer.
DOC? <sp>mm/dd/yyyy<cr><lf> Returns the date of cal for the active
transducer.
DOM? <sp>mm/dd/yyyy<cr><lf> Returns the date of manufacture
Error? <sp> text description <cr><lf> Returns the next error in the error
queue
Errorno? <sp>Enn-text<cr><lf> Returns pcs400 error code and text
Filter OFF, LOW, NORMAL, HIGH Sets the reading lter 0, 80%, 92%,
95%
Filter? <sp> (lter)<cr><lf> Returns the reading lter.
Gasdensity Value in lb/cuft, or “NITROGEN” or
“DRYAIR”
Sets the head pressure gas density in
lb/cuft
Gasdensity? <sp>+n.nnnnnE+nn<cr><lf> Gets the head pressure gas density in
lb/cuft
Gastemp Value in degrees F Sets the head pressure gas tempera-
ture
Gastemp? <sp>+n.nnnnnE+nn<cr><lf> Gets the head pressure gas tempera-
ture
Gateway nnn.nnn.nnn.nnn Sets the Ethernet gateway address
Gateway? <sp>nnn.nnn.nnn.nnn<cr><lf> Gets the Ethernet gateway address
Height Value in inches Sets the head pressure height in
inches
Height? <sp>+n.nnnnnE+nn<cr><lf> Gets the head pressure height in
inches