Operating Instructions - CPC6050 79
Modular Pressure Controller
CPC6050
Highspeed ON or OFF Sets the highspeed control mode,
which sets the control behavior, stable
window, stable delay, and rate set-
point
Highspeed? <sp>(YES or NO)<cr><lf> Returns the status of the highspeed
mode
Id? <sp>MENSOR,CPC6050,
ssssss,v.v.vv<cr><lf>
Ssssss is the serial number,v.v.vv is
the CPC6050 software version.
Instrument_control_
time_on?
<sp>NNNNN<cr><lf> Returns the amount of time the instru-
ment has been in control mode since
the last reset of the counter, in min-
utes
Instrument_total_con-
trol_time_on?
<sp>NNNNN<cr><lf> Returns the lifetime amount of time
the instrument has been in control
mode, in minutes
Instrument_time_on? <sp>NNNNN<cr><lf> Returns the amount of time the instru-
ment has been on since the last reset
of the counter, in minutes
Instrument_total_
time_on?
<sp>NNNNN<cr><lf> Returns the lifetime amount of time
the instrument has been on, in min-
utes
IP nnn.nnn.nnn.nnn Sets the IP address of the instrument
IP? <sp>nnn.nnn.nnn.nnn<cr><lf> Returns the IP address of the instru-
ment
Keylock YES or NO Locks or unlocks the entire touch
screen
Keylock? <sp>(YES or NO)<cr><lf> Returns Yes or No
Leak_Chan A, B Sets the leak channel over the bus
and makes that channel the active
remote channel
Leak_Chan? <sp>(A or B)<cr><lf> Returns the active leak channel
Leak_External ON or OFF Sets the leak test external valve set-
ting in Single output A or B mode only
Leak_External? <sp>(YES or NO)<cr><lf> Returns the leak test external valve
setting
Leak_Start Changes the screen to the Leak Test
if on another screen, begins the leak
test with current parameters.
Leak_Stop Stops the test and vents the system.
Leak_State? <sp>(IDLE or RUNNING or
COMPLETE)<cr><lf>
Returns state of the leak test
Leak_Setpt Value inside upper and lower limits Starting point for the leak test to begin
at
Leak_Setpt? <sp>n.nnnnnE+nn<cr><lf> Returns starting setpt
Leak_Dwell 0 to 3600 The duration of the test during which
a leak is measured