80 Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
Leak_Dwell? <sp>n.nnnnnE+nn<cr><lf> Returns the dwell time in seconds
Leak_Pre_Dwell 0 to 3600 Pre-dwell parameter is the time in
between going to measure mode, and
the dwell time. Useful if there are
pressure transients after switching the
measure.
Leak_Pre_Dwell? <sp>n.nnnnnE+nn<cr><lf> Returns the pre-dwell time.
Leak_Delta_Limit Value within the range of the sensor Threshold for a passing/failing leak
test.
Leak_Delta_Limit? <sp>n.nnnnnE+nn<cr><lf> Returns the maximum pressure
change allowed before failure
Leak_Pass? <sp>(YES or NO)<cr><lf>
Leak_Initial? <sp>+n.nnnnnE+nn<cr> Returns the leak test initial pressure
Leak_Final? <sp>+n.nnnnnE+nn<cr> Returns the leak test nal pressure
Leak_Delta? <sp>n.nnnnnE+nn<cr><lf> Returns the change in pressure since
start of test.
List? <sp>Pri,1,2;Sec,1,2;Bar,1<cr><lf> Returns list of available transducers
and turn downs for the active channel.
Listcal? <sp>PRI,{sn},1,{mm
ddyyyy},2,{mmdd
yyyy};SEC,{sn},1,{mmdd
yyyy},2,{mmddyyyy}<cr>
<lf>
Returns the serial number of each
installed transducer and calibration
dates for each range
Listrange? <sp>PRI,1,min,max,2,min,max;S
EC,1, min,max,2,min,max[;3RD,1,min,
max,2,min,max;4TH,1,min,max,2,min,
max]<cr><lf>
(bracketed single output only)
Returns the ranges of the installed
transducers for the active channel.
Listsensorinfo? <sp>PRI,SN,PTYPE,ACCURACY,1,
min,max,2,min,max;SEC,SN,PTYPE,
ACCURACY,1,min,max,2,min,max[;3
RD,SN,PTYPE,ACCURACY,1,min,m
ax,2,min,max;4TH,SN,PTYPE,ACCU
RACY,1,min,max,2,min,max]<cr><lf>
(bracketed single output only)
Returns the serial number, pressure
type, accuracy, and ranges of the in-
stalled sensors for the active channel
Listcong? CHA, Regtype;PRI,1,min,max;SE
C,1,min,max; CHB, Regtype;PRI,1
,min,max;SEC,1,min,max; CHbaro,
Baro;Bar,min,max
Returns the regulator types (Regtype:
Pump, LPSVR, MPSVR, HPSVR,
EPSVR) along with all the ranges of
the installed transducers.
Localgravity Value in ft/s^2 Sets the local gravity in feet/sec^2
Localgravity? <sp>+n.nnnnnE+nn<cr><lf> Returns the local gravity in feet/sec^2
LowerLimit Value inside primary xducer range in
current units.
Sets the lower control limit for the
instrument
LowerLimit? <sp>n.nnnnnE+nn<cr><lf> Returns the lower control limit for the
instrument in current units.
Macaddress? <sp>cc-cc-cc-cc-cc-cc<cr><lf> Returns Ethernet MAC address
Measure None Instrument placed in Measure Mode