EasyManuals Logo

mensor CPC6050 User Manual

mensor CPC6050
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Operating Instructions - CPC6050 77
Modular Pressure Controller
CPC6050
Calibration_mode YES, NO Enable/ disable the calibration mode,
which when on, disables the sensor
overpressure error and forces range
hold to lock the secondary shuto
valve. Normally the secondary range
is always allowed to go to its maxi-
mum pressure to avoid a large pres-
sure disturbance when switching to it.
Cerr None Clears the error queue
Chan A, B, D or Baro Select the current channel for remote
communication. Does not apply to
single output autorange option. For
single output channel A or B, this
selects the current channel. Channel
Baro should only be set when used
for calibration functionality
Chan? <sp>A,B,D, or BARO<cr><lf> Return the current channel for remote
communication
Cmdset Mensor, DPI510, SCPI Activates remote command set for
instrument emulation modes.
Cmdset? <sp><CCCCCC><cr><lf> Returns active command set identier
Control Instrument placed in Control Mode
Control? <sp>(YES or NO)<cr><lf> Returns YES if instrument is in control.
NO if otherwise
Control_behavior <n> 0 to 100 100 = highspeed, 0 = low overshoot.
May be set between 0 and 100.
Control_behavior? <sp>NNN<cr><lf> Retuns the control behavior number
Control_default PRECISION, HIGHSPEED, CUSTOM
or NONE
Sets the current control settings
Control_default? <sp>(PRECISION, HIGHSPEED,
CUSTOM or NONE)<cr><lf>
Returns the current control settings
Crate Slow, Medium, Fast, Variable Sets the control rate, Variable mode
is a predetermined user dened rate
setpt.
Crate? <sp>CCCCCC<cr><lf> Returns the control rate – CCCC is
variable in length and corresponds to
the parameters for the CRATE com-
mand
Ctype? Returns the type of regulator
Decpt? <sp>n<cr><lf> Returns the number of decimal points
(see Resolution)
Default None Sets the default values
Deltafunc A+B, A-B, B-A Sets the Delta channel’s function
Deltafunc? <sp>{A+B or A-B or B-A}<cr><lf> Returns the Delta channel’s function
DHCP ON or OFF Reserved for DHCP setup
DHCP? <sp>(YES or NO)<cr><lf> Reserved for DHCP setup

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the mensor CPC6050 and is the answer not in the manual?

mensor CPC6050 Specifications

General IconGeneral
Brandmensor
ModelCPC6050
CategoryController
LanguageEnglish

Related product manuals