Operating Instructions - CPC6050 85
Modular Pressure Controller
CPC6050
Switch_Chan? <sp>(A or B)<cr><lf> Returns the active switch channel
Switch_Start Changes the screen to the Switch
Test if on another screen, begins the
switch test with current parameters.
Switch_Stop Stops the test and vents the system.
Switch_State? <sp>(IDLE or RUNNING or
COMPLETE)<cr><lf>
Returns state of the switch test
Switch_Low Value inside upper and lower limits Starting setpt for the switch test to
begin at
Switch_Low? <sp>+n.nnnnnE+nn<cr><lf> Returns starting setpt
Switch_High Value inside upper and lower limits Finishing setpt for the switch test to
end at
Switch_High? <sp>+n.nnnnnE+nn<cr><lf> Returns ending setpt
Switch_Fast Value inside upper and lower limits Rate at which switch test will go to the
low setpt
Switch_Fast? <sp>+n.nnnnnE+nn<cr><lf> Returns fast rate setpt
Switch_Slow Value inside upper and lower limits Rate at which switch test will go to the
high setpt
Switch_Slow? <sp>+n.nnnnnE+nn<cr><lf> Returns slow rate setpt
Switch_1_Up? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
1 tripped while going up scale. Re-
turns 0 if switch was not triggered.
Switch_1_Dn? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
1 tripped while going down scale. Re-
turns 0 if switch was not triggered.
Switch_2_Up? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
2 tripped while going up scale. Re-
turns 0 if switch was not triggered.
Switch_2_Dn? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
2 tripped while going down scale. Re-
turns 0 if switch was not triggered.
Switch_3_Up? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
3 tripped while going up scale. Re-
turns 0 if switch was not triggered.
Switch_3_Dn? <sp>+n.nnnnnE+nn<cr><lf> Returns the pressure at which switch
3 tripped while going down scale.
Ta r e ON / OFF Tares the reading to zero
Tare? <sp> n.nnnnnE+nn <cr><lf> Returns value of Tare
Termchar CCCC Set the ouptput termination
character(s). CR, LF, CRLF, DEFAULT,
EOI, NONE
Termchar? Returns the termination character
setting
Transfer_factory_to_
linearity
Copy factory linearity coecients to
customer