Programming
142
Duration: 0. 15 to 59 Sec.
0 to 99 min.
Rhythm
No
Normal
Delicate
Gentle
User-defined:
On time 2 to 99Sec.
Off time 2 to 99Sec.
Speed
No
Normal
Scooping
User-defined 20 to 70rpm.
17. Thermostop
Thermostop is locked when "No" is programmed in program
stage 4. Heating
Choose from the following options:
– No
When Thermostop "No" is set, the heating phase begins when the
minimum heating level is reached. Wash times 1and2 are completed
chronologically without necessarily having to reach the temperature
selected.
– Yes
When Thermostop "Yes" is programmed the heating phase also be-
gins when the minimum heating level is reached and then allows
Wash time 2 to run only when the target temperature is reached.
Wash time 2 always starts at the target temperature.
– Disinfection
When Thermostop "Disinfection" is programmed the procedure is the
same as for "Yes". Additionally, after a drop in temperature of 3°C
below the lowest permissible hysteresis value (e.g. in case of peak
load, refueling, power cut), Thermostop is activated and Wash time2
and Dispensing phase2 have to run again.
18. Level 2
Settings as for Level 1
19. Intake path 2
Settings as for Intake path 1
20. Dispensing
phase 2
Settings as for Dispensing phase 1