Miele PW 811 - Cleaning and Maintenance of the Dispensing System

Miele PW 811
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Optional modules
45
Dispensing calibration
Dispensing pump 1
0/1
0 st:
P: No
V: 100 ml
Press multifunction button 0/1
Stop the pump by pressing the 0/1 button, or it will stop automatically
after 60seconds.
Measure the amount of liquid dispensed into the measuring vessel.
Dispensing calibration
Dispensing pump 1
0/1
30 st: P: 200 ml/minV: 100 ml
Enter the measured value into the"V:" field.
The throughput P is calculated in ml/min. by the controls.
Press the "Save" button.
The measured value is allocated to the pump.
Reconnect the hose.
Cleaning and maintenance of the dispensing system
Rinse the dispensing system through with hot water every 2 weeks,
or before shutting the dispenser pumps down for any length of time,
to avoid the system getting clogged up and corroded.
To do this, rinse out the siphons with water.
Put the siphons in a container with warm water (104 - 122°F / 40 -
50°C).
Activate the dispenser calibration switch in each pump, until the
dispensing system is rinsed through.
Check the connections, hoses, dispensing sockets and seals for
leaks.