XII
2. Confirm that the probe and cable are normal before and
after each examination. A defective probe may cause
electric shock to the patient.
3. Do not subject the probe to shock. A defective probe may
cause electric shock to the patient.
4. Do not disassemble the probe to avoid the possibility of
electric shock.
5. Never immerse the probe connector into liquids such as
water or disinfectant because the connector is not
waterproof. Immersion may cause electric shock or
malfunction.
6. Ultrasonic probe is only for use with the specified
ultrasonic diagnostic system. Please refer the ultrasonic
diagnostic system operation manual to select the proper
probe.
7. A probe sheath must be installed over the probe before
performing intra-cavity or intra-operative examination.
CAUTION:
1. When using the probe, wear sterile gloves to prevent
infection.
2. Be sure to use ultrasound gel. Please use the ultrasound gel
compliant with the relevant local regulations.
3. In normal diagnostic ultrasound mode, there is no danger of a
normal-temperature burn; however, keeping the probe on the
same region of the patient for a long time may cause such a
burn.
4. Do not use the carrying case for storing the probe. If the
carrying case is used for storage, it may become a source of
infection.
5. The probe and accessories supplied with it are not delivered
disinfected or sterilized. Sterilization (or high-level disinfect)
before use is required.
6. It is required to practice ALARA when operating ultrasound
system. Minimize the acoustic power without compromising
the quality of images.
7. Disposable components are packaged sterile and are
single-use only. Do not use if integrity of packaging violated
or if expiration date has passed. Please use the disposable
components compliant with the relevant local regulations.
8. Please use the disinfection or sterilization solution that
recommended in this operator’s manual, otherwise Mindray
will not be liable for damage caused by other solutions. If you
have any questions, please contact Mindray Customer Service
Department or sales representative.
9. Do not use pre-lubricated condoms as a sheath. Lubricant
may not be compatible with the transducer material and
damage may result.