EasyManuals Logo

Mindray M5 User Manual

Mindray M5
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #271 background imageLoading...
Page #271 background image
DICOM
B-2
Local TCP/IP setting items are described as follows:
1 Station Name: Name of the ultrasound system
2 Station Location: Location or department where the system is placed.
3 Current Net Adapter: to select the network connection mode
4 DHCP/ Static: If “DHCP” is selected, IP address will be automatically obtained from
DNS server; if “Static” is selected (using static IP address), you need to enter the IP
address.
5 IP Address: IP address of the system; should be at the same network segment with
the service IP address.
6 Subnet Mask: Used to set different network segment.
7 Gateway: Used to set the gateway IP.
8 OK: Click to confirm parameter setting.
9 Cancel: Click to cancel parameter setting.
Tips:
z The IP address of the system should not be occupied by other devices in the LAN,
or failure occurs.
z After you changed “Station Name”, you need to restart the system to make the
setting effective (press the power button and select “Shut Down” and restart the
system after a while).
B.1.2 DICOM Local Setting
To set the DICOM server properties and log saving manner.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray M5 and is the answer not in the manual?

Mindray M5 Specifications

General IconGeneral
BrandMindray
ModelM5
CategoryMedical Equipment
LanguageEnglish

Related product manuals