DICOM
B-2
Local TCP/IP setting items are described as follows:
1 Station Name: Name of the ultrasound system
2 Station Location: Location or department where the system is placed.
3 Current Net Adapter: to select the network connection mode
4 DHCP/ Static: If “DHCP” is selected, IP address will be automatically obtained from
DNS server; if “Static” is selected (using static IP address), you need to enter the IP
address.
5 IP Address: IP address of the system; should be at the same network segment with
the service IP address.
6 Subnet Mask: Used to set different network segment.
7 Gateway: Used to set the gateway IP.
8 OK: Click to confirm parameter setting.
9 Cancel: Click to cancel parameter setting.
Tips:
z The IP address of the system should not be occupied by other devices in the LAN,
or failure occurs.
z After you changed “Station Name”, you need to restart the system to make the
setting effective (press the power button and select “Shut Down” and restart the
system after a while).
B.1.2 DICOM Local Setting
To set the DICOM server properties and log saving manner.