Refers to time after which the system will stop
trying to establish a connection to the service.
Value: 5-60s, in increments of 5, and 15 by
default.
This function enables the ultrasound system to
use the listen port for data receiving.
Port for ultrasound system to receive data after
the listen mode function is activated. Here, the
port should be consistent with that of the
Worklist server port.
For details of listen port setting, refer to settings
in the server.
Click to add the Worklist service to the service
list.
Click to cancel the parameter setting.
Select an item in the service list, change the
parameters in the above area, and click
[Update] to update the item in the service list.
Click to delete the selected service in the
service list.
Select an item in the service list, click [Default]
and you can see “Y” in the Default column.
OTHERS
You can configure different type of DICOM services in different scenarios.
DICOM Verifying
To verify connectivity (not essential), click the [Verify] button on the Storage, Print,
Worklist, MPPS, Storage Commitment and Query/Retrievepages respectively.
If the verification is successful, the system displays “xxx Verification Succeed.”
Otherwise, it displays “xxx Verification Failed.”
If verification failed, possible causes may be: wrong IP address; not able to access IP
address; remote DICOM server is not running; wrong port; incorrect application (AE)
Title.
Tip:
Not all SCPs can support verification. See the SCP properties to confirm whether the
SCP can support this service. If not, the verification will not be successful.