3 System Specifications
This section contains system and accessory specifications. For information on the
specifications for Mindray/Zonare authorized peripherals, refer to the manufacturers’
documentation.
General System Specifications
Type
Parameter
Value
Electrical
Power requirements
100-240V~, 50-60Hz, 6A max
Power consumption (no
peripherals max)
180W (616 BTU/hr)
Power consumption (with
peripherals max)
470W (1608 BTU/hr)
Environmental
(operating)
Cooling requirements
See power consumption above
Air temperature
0-35°C (32 95F)
Humidity
15-80% noncondensing
Pressure
700-1060 hPa
Environmental
(storage)
Air temperature
-20-60°C (-4-140°F)
Humidity
15-90%, noncondensing
Pressure
500-1060 hPa
Physical
(weight)
ZS3 system
91.5 kg (without probes)
Power cord
0.6 kg (1.4 lbs.)
USB printer & bracket
3.4 kg (7.6 lbs.)
Battery & bracket
4.1 kg (9.1 lbs.)
Operating weight subtotal
73.4 kg (162.1 lbs.)
System shipping
container
41.5 kg (91.6 lbs.)
Total shipping weight
114.9 kg (253.7 lbs.)
Physical
(dimensions)
Height, max (in operating
use)
151 cm (59.4 in.)
Height, min (in operating
use)
131 cm (51.5 in.)
Height, min (display
lowered for transport)
106 cm (41.7 in.)

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals