EasyManuals Logo

Mindray Zonare ZS3 Service Manual

Mindray Zonare ZS3
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #109 background imageLoading...
Page #109 background image
ZS3 Service Manual
Page 109 of 295
3. Select the destination to “DICOM Media” and DICOM compression mode. You can
select to delete the exam or the image after the backup.
4. Click [OK]. The image from the current exam is sent to the external storage media in
DICOM format.
If the backup is successful, a tick will appear in the Backup list in the Archive screen. If
not, there will be no tick.
There must be no DICOMDIR/DCMIMG/IHE_PDI files on the external storage media of
the same name as the one being backed up. Otherwise, the backup cannot proceed.
Ensure there is enough storage space, or the backup may fail due to shortage of space.
Media review
1. Connect the external media with DCM files to the system.
2. Select the data source in Archive screen, and the visible data will be shown.
If there are several types of data on the media, the system will ask you to select the
format. Then, click [DICOMDIR].
Data Restore
After the DICOM format data are saved to external media, restore the data to the
ultrasound system.
1. Connect the external media containing DCM files to the system.
2. In Archive, review the data stored on the external media.
3. Select the data to be restored in Archive.
4. Click on the Archive screen.
NOTE:
Only system-accessible media can be selected.

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals