DP551U. This is a plug-and-play wireless networking solution that is compatible with all
Mindray/Zonare ultrasound systems that support Ethernet connectivity. A special power
cable, however, is required to supply power to the Quatech AirborneDirect device
from a USB port on the Mindray/Zonare system, for portable operations.
The Quatech AirborneDirect Wireless Ethernet Bridge supports WiFi 802.11b/g
wireless standards and connects to the Mindray/Zonare system through a 10 Base-T
network interface. This device supports WEP (64/128 bit) and WPA encryption
standards, and LEAP for network authentication (LEAP requires that Quatech device be
configured with a static 128 bit key; this is known as "migration mode" and is not
recommended for long term use due to the static WEP key requirement). Before using it
with the Mindray/Zonare system, the Quatech AirborneDirect device must be
preconfigured for the user's networking environment – SSID, channel, encryption, and
correct addressing scheme (DHCP/static).
Static IP addressing is required in order to be able to display the additional wireless
status. When power is applied to a correctly configured Quatech AirborneDirect
device that is connected to the Ethernet port on the Mindray/Zonare system, it will
require 30 to 60 seconds to detect and then associate with the user's wireless network.
Once a wireless network connection is established, the Mindray/Zonare system will
reflect the connection state by showing the network ICON on screen, in an uncrossed
state.
The configuration of the Wireless Bridge is done on the Network configuration screen.
The source information required for wireless device configuration on the ZS3, is the
static IP Address assigned (by the IT Administrator at the install site) to the Quatech
AirborneDirect device and the Username and Password for the CLI interface (default
is dpac for both).
For assistance with the configuration of the local network settings on the ZS3, to
support that aspect of a wireless network installation, contact Mindray/Zonare Technical
Support.
Wireless bridge settings
1. Connect the Airborne Direct wireless bridge to the network port and the power
cable to the USB port of the ultrasound system for the configuration and ensure
the power has been connected to it.
2. Enters [Setup] [Network] [TCP/IP].