current leakage and isolation from a primary power line. Testing for compliance with the
essential requirements of the Medical Device Directive has been performed.
Warnings
Do not remove any of the System covers other than the module cover. Other
than the scan module there are no user-serviceable parts internal to the
system. Only trained Mindray/Zonare service personnel should access the
system’s internal electronics.
The ultrasound systems contain no operator-serviceable components within
the enclosures. To avoid electrical shock, do not remove covers. As with any
other electrical equipment, always observe care when operating this
instrument. For service issues, contact Mindray/Zonare Technical Support.
Failure to follow these restrictions may void your warranty or service contract
coverage.
To reduce the risk of electric shock, DO NOT connect the ZS3 system input
or output connections to equipment that is not properly connected to an Earth
ground.
To achieve proper grounding reliability, the ultrasound system power plug
must be fully inserted into a receptacle marked “hospital grade.” Do not
remove the grounding wire. If there is any question of power outlet or power
cord integrity, do not proceed. Obtain qualified assistance
To maintain proper grounding reliability, use only Mindray/Zonare-
recommended peripherals and accessories. Use of non-specified peripherals
and accessories could result in risk of electrical shock or injury.
The ultrasound systems represent a potential explosion hazard if used in the
presence of flammable anesthetics.
The system does not contain a user-serviceable lithium ion battery.
The optional ZPAK cart battery is not a user serviceable item. Contact
Mindray/Zonare’s Technical Support group for assistance with the ZPAK
battery.
Follow guidelines provided by relevant IEC standards) when connecting
peripherals.
The USB Memory Sticks supplied by Mindray/Zonare are the recommended
brand, type, and sizes for use in ZS3 Systems. They have been verified for
optimum reliability and performance.
No modification of the system is permitted. Modifying the system may subject
the operator or patient to hazardous conditions.
The ZS3 ultrasound system represents a potential explosion hazard if used in
the presence of flammable gases or oxygen rich environment.
Use only transducers that are specifically approved for the ultrasound system.
If the proper identification of a connected transducer is not displayed on
screen, do not proceed with its use.
Transducers covers may be contaminated and must be handled accordingly.