Error Code
Subsystem
Message for Error log
User Error Message
Action on
Action
0xe00a042e
IMAGING
Failed to do CPLD sync
"Internal System Error
has occurred. "
Restart
0xe00a042f
IMAGING
HV1N/HV2N/HVCW
comparison error (bad
command inputs)
"Internal System Error
has occurred. "
Restart
0xe00a0430
IMAGING
HV limits check failed (bad
command inputs)
"Internal System Error
has occurred. "
Restart
0xe00a0431
IMAGING
Failed to set DACs
(communication error)
"Internal System Error
has occurred. "
Restart
0xe00a0432
IMAGING
HV's aren't in regulation
"Internal System Error
has occurred. "
Restart
0xe00a0433
IMAGING
Command not allowed in
present state
"Internal System Error
has occurred. "
Restart
0xe00a0434
IMAGING
Command not allowed in
error state
"Internal System Error
has occurred. "
Restart
0xe00a0435
IMAGING
HV Alert unknown Error
"Internal System Error
has occurred. "
Restart
0xe00a0436
IMAGING
Received PFE Fault
"Internal System Error
has occurred. "
Restart
0xe00a0437
IMAGING
Check HVInput - invalid Input
parameters, see log
statement above
"Internal System Error
has occurred. "
Restart
0xe00a0439
IMAGING
Timeout out waiting for In
Regulation message after
setting HV's
"Internal System Error
has occurred. "
Restart
0xe00a043a
IMAGING
PowerMicro timed-out
waiting for HV in regulation
"Internal System Error
has occurred. "
Restart
0xef050402
THERMALSER
VICES
Thermal Control Sensor
Error
"Thermal Sensor Error:
shutting down system."
Shutdown
0xef050404
THERMALSER
VICES
Exceeded maximum system
error count
"System Temperature
Limit exceeded."
Shutdown
0xef050406
THERMALSER
VICES
Battery temperature error
"Battery Temperature
Limit Exceeded. Replace
battery."
Shutdown
0xef050420
THERMALSER
VICES
PWR_HV over Temperature
"System Temperature
Limit exceeded."
Shutdown
0xef050421
THERMALSER
VICES
PWR_LV over Temperature
"System Temperature
Limit exceeded."
Shutdown
0xef050422
THERMALSER
VICES
PWR_TSUPR over
Temperature
"System Temperature
Limit exceeded."
Shutdown

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals