Page #211 background imageLoading...
Page #211 background image
ZS3 Service Manual Page 211 of 295
Error Code
Subsystem
Message for Error log
User Error Message
Action on
Action
0xef050466
THERMALSER
VICES
XMT_VSUPR sensor
communication error
"System Temperature
Sensor Communication
Error."
Shutdown
0xef050468
THERMALSER
VICES
RCV_AFE sensor
communication error
"System Temperature
Sensor Communication
Error."
Shutdown
0xef050469
THERMALSER
VICES
RCV_CW sensor
communication error
"System Temperature
Sensor Communication
Error."
Shutdown
0xef05046a
THERMALSER
VICES
RCV_LV sensor
communication error
"System Temperature
Sensor Communication
Error."
Shutdown
0xef05046b
THERMALSER
VICES
RCV_VSUPR sensor
communication error
"System Temperature
Sensor Communication
Error."
Shutdown
0xef05046c
THERMALSER
VICES
Multiple Temperature Sensor
Error
"System Temperature
Sensor Failure."
Shutdown
0xef05046d
THERMALSER
VICES
Temp Sensor Alert received
with unknown error
"System Temperature
Sensor Failure."
Shutdown
0xef050470
THERMALSER
VICES
Failed read alert status
"Temperature Limit
Failure."
Shutdown
0xef050480
THERMALSER
VICES
Fan is stuck
"System fan Error:
Shutting down."
Shutdown
0xef070401
DISPLAY
DVI configuration failed
"Monitor Configuration
Error: Please shutdown
system."
Shutdown
0xef080401
FRONT_END_P
OWER
Low Voltage failure occurred
"Internal System Error
has occurred. "
Shutdown
0xef080402
FRONT_END_P
OWER
Relay Voltage failure
occurred
"Internal System Error
has occurred. "
Shutdown
0xef080403
FRONT_END_P
OWER
3.3 Voltage failure occurred
"Internal System Error
has occurred. "
Shutdown
0xef080404
FRONT_END_P
OWER
V5_D Voltage failure
occurred
"Internal System Error
has occurred. "
Shutdown
0xef0e0401
ROUTER
Video Memory Test Failed
"Internal System Error
has occurred. "
Shutdown
0xef0e0402
ROUTER
DCM LOCK Fault has
occurred
"Internal System Error
has occurred. "
Restart

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals