EasyManuals Logo

Mindray Zonare ZS3 Service Manual

Mindray Zonare ZS3
295 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #275 background imageLoading...
Page #275 background image
ZS3 Service Manual Page 275 of 295
Appendix A Phantom Usage
Illustration
Targets Disposal of Phantom KS107BD
A1——A5 Axial resolution target group
B: Blind-area target group
C Longitudinal target group
D Horizontal target group
E: Mimic tumor
F Mimic sac (diam 10mm) and stone
G Mimic sac (diam 6mm)
4. Line Target System
There are 8 groups of nylon line targets disposed as shown in the figure.
1A1——A5
Axial and lateral resolution target group. The distances between the horizontal branch
and the acoustic window are 30, 50, 70, 120 and 160mm, the center horizontal
distances between two adjacent lines of A1 and A2 groups are 1, 5, 4, 3, 2mm, A3~A5
groups are 5, 4, 3, 2mm. The center longitudinal distances between two adjacent lines
of the longitudinal branches are 4, 3, 2, 1mm.
2B
Blind-area target group. Center horizontal distance of adjacent lines is 10mm,
distances to the acoustic window are 10, 9, 8, 7, 6, 5, 4, 3mm.
3C
Longitudinal target group. 19 target lines with a 10mm center distance between
adjacent lines.
4D
Horizontal target group. 7 target lines with a 20mm center distance between adjacent lines.

Table of Contents

Other manuals for Mindray Zonare ZS3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Zonare ZS3 and is the answer not in the manual?

Mindray Zonare ZS3 Specifications

General IconGeneral
BrandMindray
ModelZonare ZS3
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals