ELECTRICAL SAFETY INSPECTION
2- Device Enclosure And Accessories
TEST PROCEDURE
Visual Inspection
The Enclosure and
Accessories
No physical damage to the
enclosure and accessories.
No physical damage to meters,
switches, connectors, etc.
No residue of fluid spillage (e.g.,
water, coffee, chemicals, etc.).
No physical damage to probe
head (e.g., crack)
No loose or missing parts (e.g.,
knobs, dials, terminals, etc.).
Contextual Inspection
The Enclosure and
Accessories
No unusual noises (e.g., a rattle
inside the case).
No unusual smells (e.g., burning
or smoky smells, particularly from
ventilation holes).
No taped notes that may suggest
device deficiencies or operator
concerns.