ZS3 Service Manual Page 290 of 295
ELECTRICAL SAFETY INSPECTION
7- Mains on Applied Part Leakage
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if
situation cannot be corrected, submit a Safety Failure Report to document
the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for
analysis and disposal.
LIMITS
All countries:
For BF ECG input and transducer:
5000μA