If the current test standard being used does not include Patient Auxiliary
Current DC readings, or the DC option is not enabled, then DC readings will not be
available through the APPLIED PART SOFT KEY selections.
Check any broken of the enclosure. Replace any defective part.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other wiring
abnormalities. Notify the user or owner to correct any deviations. As a work
around, check the other outlets to see if they could be used instead.
Change another probe to confirm if the fail is caused by console.
Inspect wiring for bad crimps, poor connections, or damage.
If the leakage current measurement tests fail on a new unit and if
situation cannot be corrected, submit a Safety Failure Report to document
the system problem. Remove unit from operation.
If all else fails, stop using and inform the Customer Service Engineer for
analysis and disposal.