IEC 60601-1-2: 2014 Certification
Specification
Compliance
IEC 61000-4-5 surge
Compliance level:
±1 kV differential mode
±2 kV common mode
Electromagnetic environment guidance:
Main power quality should be that of a typical commercial or
hospital environment.
IEC 61000-4-6 Conducted Immunity
Compliance level:
0.15MHz - 80 MHz
3Vrms @ 2Hz
IEC 61000-4-8 power frequency (50/60
Hz) magnetic field immunity
Compliance level:
3A/m, 50/60Hz
Electromagnetic environment guidance:
Power frequency magnetic fields should be at levels
characteristic of a typical commercial or hospital
environment.
IEC 61000-4-11 voltage dips, short
interruptions, and voltage variations on
power supply input lines
Electromagnetic environment guidance:
Main power quality should be that of a typical commercial or
hospital environment.