Assign
from
Modality
Worklist
In the worklist, search for exams with selected modalities.
Verify the effectiveness of the TLS certificates.
Import trusted certificates.
Application Entity title.
Communication port, DICOM communication port. The general port
of DICOM port is 2345 by default.
Maximum PDU data package size (not need to change), ranging
from 16384 to 65536; if the value is less than 16384 or greater than
65536, the system automatically sets it to the value 32768.
Name of the device supporting DICOM services.
IP address of the server.
You can ping the other machines after you entered the correct IP
address.
Besides, you can select a server in the Device list to ping it.
Displays the added device.
Provides server settings of DICOM service, for details, please refer
to the following chapters.
Click to add server (s) to the Device List.
Click to delete the selected server (s) in the device list.
Server setting procedure:
a) Enter the server device name and IP address; click [Ping] to check the connection.
b) Click [Add] to add the server to the device list, and its name and address are
displayed in the Device List.
Tips:
AE Title should be the same with the SCU AE Title preset in the server
(PACS/RIS/HIS), for example, if the AE Title of the server preset in the storage
server is Storage, and the AE Title of the accepted SCU is preset as Machine, then
in the figure above, the AE Title of Local should be Machine, and the AE Title of
storage server should be Storage.
The device name can be random. If the server name entered in the Device field
matches that of another Device in the DICOM Device List, the following message
will appear, “the server added already exits.” Click [Ok] to enter a different
Server/Device name.