Chapter 3: Installation 3-21
Aftercalibratingthescale,theaccuracymustbeverifiedtoassureitmeetsthe
governmentrequirements.Thisconsistsofaseriesoffourteststhatmustberuninthe
continuoussequencegiven.
1. IncreasingLoadTest
2. Over‐CapacityTest
3. DecreasingLoadTest
4. ShiftTest
Increasing Load Test
Thistestchecksthescale’saccuracywhenincrementallyaddingweighttothecenter
oftheTopPlate.UseweightsthatcorrespondtotheRealScan74weightfeatures.
24177
Step Weight Feature Add Weight Remove Weight Display Result
1
2
3
4
5
Kilogram 0.100
kg
Pound 0.20 lb
Pound
2.500 kg
5.00 lb
0.10
kg
0.20 lb
Pound
5.000 kg
5.00 lb
Pound
5.000
kg
10.00 lb
Pound
2.500 kg
10.00 lb
0.100 0.000 kg
+
0.20 0.00 lb
+
Kilogram
Kilogram
Kilogram
Kilogram
2.500 0.000 kg
+
5.00 0.00 lb
+
7.500 0.005 kg
+
10.00 0.01 lb
+
15.000 0.005 kg
+
12.500 0.005 kg
+
20.00 0.01 lb
+
30.00 0.01 lb
+
Note: DoNOTremoveanyweightfromtheTopPlatebetweenteststepsunless
directedtodoso.
Over-Capacity Test
Thistestchecksfortheproperindicationfromthescalewhentoomuchweightis
placedontheTopPlate.
Note: ThistestmustimmediatelyfollowtheIncreasingLoadTest;donotremove
anyoftheweightsfromtheprevioustestpriortorunningthistest.
Note: Scalecapacityis30lb
or15kgin0.01lbor0.005kgincrements.