7.1 Error messages
FRA5087
No. Messages Descriptions
----------------------------------------
13 Can't load, file corrupted The file is corrupted or is not correct, and
therefore it cannot be loaded.
14 Can't save during acquisition Data is still in acquisition phase, and therefore
they cannot be saved.
15 Amplitude compression failure Designated amplitude could not be set. Refer to
oscillator maximum output level and amplitude
variations, and review settings.
16 Storage area overflow Memory capacity is not sufficient for storage and
overflow has occurred. Measurement data
cannot be stored anymore.
17 Can't set, range is too narrow The sweep range which you are trying to set is
narrower than the frequency resolution capability.
You may not set this sweep range.
18 Can't set, line markers are invalid
Since the line markers in the y-axis direction are
invalid for the sweep range that you are setting,
you may not set the sweep range by using the
line markers.
19 Can't set, markers aren't displayed
Since markers are not displayed, you may not set
what you are trying to do.
20 Can't measure, analysis frequency is over limit
Analysis frequency is over 10 MHz which is the
upper limit, and therefore, measurement cannot
be done.
21 Can't set, line markers are invalid
Since the line markers in the y-axis direction are
invalid for the sweep range that you are setting,
you may not set the sweep range by using the
line markers.
23 Missing file name The file name is missing.
24 Duplicate file name Duplicated file names have been found.