Itala | INSTRUCTION MANUAL
5.5.3 Sensor centering data
All cameras are tested after assembly to make sure the sensor is correctly centered. Measure-
ments are taken in all six degrees of freedom relative to the lens mount to ensure optimal imaging
performance. In Table 10 and 11 are reported tipical production values for the sensor centering
relative to Figure 32.
Axis Nominal 3 σ
σ
σ
Roll 0° 0.4°
Yaw 0° 0.3°
Pitch 0° 0.5°
Horizontal shift (x) 0 mm 0.2 mm
Vertical shift (y) 0 mm 0.2 mm
FD (z) 17.53 mm 0.11 mm
Table 10: Sensor centering for C-mount cameras
Axis Nominal 3 σ
σ
σ
Roll 0° 0.6°
Yaw 0° 0.3°
Pitch 0° 0.5°
Horizontal shift (x) 0 mm 0.2 mm
Vertical shift (y) 0 mm 0.2 mm
FD (z) 12 mm 0.2 mm
Table 11: Sensor centering for J-mount cameras
v1.12 - eng 50