4.5.7 CAL Lock
CAL Lock allows the user to lock the calibration so that it cannot be
changed while CAL Lock is enabled. If the user tries to change the
calibration while the lock is activated, the gage displays a CAL Lock
message.
To activate CAL Lock:
Step 1: Press [2ndF][ ](Setup) to display the setup tabs.
Step 2: Use the [ , ] keys to highlight the MEAS tab.
Step 3: Use the [ ] key to highlight CAL Lock and [ , ] to
select between Off and On.
Step 4: Press [MEAS] to return to the Measure Mode.
4.5.8 ID Overwrite (MG2-DL Only)
ID Overwrite allows the user to lock ID locations so that once a
thickness reading is saved at the ID# location, users will be prompted
with a message asking if they wish to overwrite the stored thickness
value.
To activate ID Overwrite:
Step 1: Press [2ndF][ ](Setup) to display the setup tabs.
Step 2: Use the [ , ] keys to highlight the MEAS tab.
Step 3: Use the [ ] key to highlight ID Overwrite and [ , ] to
select between Off and On.