Page 4-38 MG2 Series
THK ONLY: Will save only the current thickness reading when
the Save key is pressed.
THK + B-scan: Will Save the Thickness and the B-scan image
when the Save key is pressed.
B-Scan Freeze Mode: Determines which thickness readings are
displayed when the [FREEZE] key is pressed during a scan.
Show Minimum: Will recall the minimum thickness in the scan to
the center of the screen, even if the minimum has scrolled off the
current B-scan screen.
Show Current: Will stop the B-scan at its current position when
the [FREEZE] key is pressed.
Max Thickness: Allows the user to set the thickness range for the B-
scan image. This should be set slightly larger than the maximum
thickness that will be encountered. The maximum value for Max
Thickness is 25.00" (635.00 mm).
4.14.1 Using the B-scan Function
To activate the B-scan feature:
Step 1: Press the [2
nd
F],[ ](SETUP) key to display the Setup
Tabs.
Step 2: Use the [ ] key to highlight the B-scan Tab.
Step 3: Use the [ ] Key to enter the B-scan Setup menu.
Step 4: Use [ and ] to select a parameter and [ and ] to
adjust the setting.
Step 5: When the Max thickness is highlighted, press [ or ]
to adjust the Max thickness (range for the B-scan) value