EasyManua.ls Logo

Panametrics MG2 - Page 71

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
March 2006 910-257C
MG2 Series Page 4-39
with the [ ,
, , ] keys and then press [ENTER].
Step 6: Press the [MEAS] key to return to the Measure mode and
save the adjusted settings.
Once the B-scan function has been enabled, the B-scan starts drawing
the cross-sectional thickness when the transducer make contact with
the material and the first thickness reading is displayed. The B-scan
will continue to draw while displaying the live thickness at the
transducer location until one of the following events occurs:
The gage encounters a LOS condition (Loss of Signal), no
thickness value. When an LOS occurs, the B-scan stops updating
and will start up again if a new thickness reading is acquired. A
vertical blank line will be inserted into the B-scan image to
indicate that a LOS condition has occurred.
The user presses the Freeze key and stops the B-scan update. The
gage will either recall the minimum thickness (Freeze Min) or
freeze at the current location (Freeze Current). This will also
display the B-scan Review screen. The user can use the [ and
] keys to review the thickness readings on the frozen B-scan.
A triangle marks the location of the minimum thickness in the B-
scan.

Table of Contents

Related product manuals